【Micro Redundant】M1K2-5AE0K2H
1500W
Server ,Workstation ,Storage System ,IPC ,Communication ,Networking
410x106x41.3(mm)
16.41x4.17x1.59(inch)
VOLTAGE: 100~264VAC FULL RANGE FREQUENCY: 47~63HZINPUT CURRENT:15A (RMS) FOR 115 VAC / 10A (RMS) FOR 230 VAC INRUSH CURRENT:15A MAX. FOR 115 VAC PER MODULE / 30A MAX. FOR 230 VAC PER MODULE(AT 25 DEGREES AMBIENT COLD START FOR EACH POWER UNIT)PFC: UP TO THE TARGET OF 95%@115V & 90% @230V, FULL LOAD.( ONE NUIT POWER SUPPLY).
* TEMPERATURE RANGE:OPERATING 0℃~50℃, STORAGE -20℃~80℃ * HOLD UP TIME:WHEN POWER SHUTDOWN ALL THE DC OUTPUT VOLTAGE MUST BE MAINTAIN 11mS IN REGULATION LIMIT AT NORMAL INPUT VOLTAGE (70% LOAD) * THE MAXIMUM POWER SUPPLY SYSTEM EFFICIENCY SHALL BE 86%, MEASURED AT NOMINAL INPUT VOLTAGE 230V AND FULL LOADING * LEAKAGE CURRENT: 3.5mA. MAX. AT NOMINAL VOLTAGE 240 VAC AND 60Hz * REMOTE ON/OFF CONTROL * OVER POWER PROTECTION:105~150% MAX * OVER VOLTAGE PROTECTION: +5V→5.7~6.5V, +3.3V→3.7~4.3V, +12V→13.2~15V * SHORT CIRCUIT PROTECTION: +5V, +3.3V, +12V, -12V, +5VSB * OVER CURRENT PROTECTION: +12V:105~150% * SAFETY: MEET60950 UL&CUL, TUV, CCC ,UL/EN/IEC 62368 * WARNING METHOD: FALUT LED, TTL * THE POWER-SUPPLY IS FOR CHASSIS-ASSEMBLY ONLY AND IS NOT ALLOWED TO BE OPERATED AS STAND-ALONE COMPONENT. FINAL ASSEMBLY HAS TO COMPLY WITH CORRESPONDING EMC- AND SAFETY-REGULATIONS.
keywords:zippy,1U Redundant,1U Single,2U Redundant,2U Single,3U Single,Cloud Power,Easy Swap,Medical,Micro Redundant,Mini 1U Single,Mini Redundant,N+1 Redundant,Open Frame,PS2,PS2*2 Redundant,PS2+,Redundant module