Products

MedicalMV1E-5350V


   

OUTPUT WATTAGE

350W

BEST SOLUTION FOR

Server ,Workstation ,PCs

DIMENSION

260x100x40.5(mm)

10.24x3.94x1.59(inch)

INPUT SPEC

VOLTAGE: 90~264VAC FULL RANGE
FREQUENCY: 47~63HZ
INPUT CURRENT:6A (RMS) FOR 115 VAC /
3A (RMS) FOR 230 VAC
INRUSH CURRENT:15A MAX. FOR 115 VAC /
30A MAX. FOR 230 VAC

DC OUTPUT SPEC

Output Voltage Output Current Min. Output Current Max. Regulation Load Regulation Line Output Ripple & Noise Max.[P-P]
+5V 0.5 20 ±5% ±1% 50mV
+12V 0.5 25 ±5% ±1% 120mV
-12V 0.1 1 ±5% ±1% 120mV
+3.3V 0.5 20 ±5% ±1% 50mV
+5VSB 0.1 2 ±5% ±1% 50mV

FEATURES

* TEMPERATURE RANGE:OPERATING 0℃~50℃, NON-OPERATING TMEPERATURE: -20℃~80℃
* HOLD UP TIME:16mS MINIMUM AT 115V LOAD <300W & 24MS MINIMUM AT 230V FULL LOAD
* EFFICIENCY:81-84% TYPICAL,AT FULL LOAD 110VAC
* POWER GOOD SIGNAL:ON DELAY 100mS TO 500mS,OFF DELAY 1Ms
* OVER POWER PROTECTION: 105~160% MAX
* OVER VOLTAGE PROTECTION: +5V→5.6V~6.5V, +3.3V→3.6V~4.3V, +12V→13.2V~15V
* SHORT CIRCUIT PROTECTION: +5V、+12V、-12V、+3.3V, +5VSB
* EMI NOISE FILTER: FCC CLASS B, CISPR22 CLASS B
* SAFETY: TO MEET UL, TUV, IEC 60601
* LEAKAGE CURRENT: 0.2mA. MAX. AT NOMINAL VOLTAGE 264VAC
* REMOTE ON/OFF CONTROL: THE POWER SUPPLY SHALL ACCEPT A LOGIC OPEN COLLECTOR LEVEL WHICH WILL DISABLE/ENABLE ALL THE OUTPUT VOLTAGE (EXCLUDE +5V STAND BY). AS LOGIC LEVEL IS LOW,OUTPUTS VOLTAGE WERE ENABLE,AS LOGIC LEVEL IS HIGH,OUTPUTS VOLTAGE WERE DISABLE.
* COOLING : ONE 40mm DC FANS

*THE POWER-SUPPLY IS FOR CHASSIS-ASSEMBLY ONLY AND IS NOT ALLOWED TO BE OPERATED AS STAND-ALONE COMPONENT. FINAL ASSEMBLY HAS TO COMPLY WITH CORRESPONDING EMC- AND SAFETY-REGULATIONS.

 

keywords:emacs,zippy,1U Redundant,1U Single,2U Redundant,2U Single,3U Single,Cloud Power,Easy Swap,Medical,Micro Redundant,Mini 1U Single,Mini Redundant,N+1 Redundant,Open Frame,PS2,PS2*2 Redundant,PS2+,Redundant module