Products

MedicalMHG2-6300P


 

OUTPUT WATTAGE

300W

BEST SOLUTION FOR

Server ,Workstation ,PCs

DIMENSION

140x150x86(mm)

5.51x5.91x3.39(inch)

INPUT SPEC

VOLTAGE: 100~240 VAC FULL RANGE
FREQUENCY : 47~63 HZ
INPUT CURRENT: 5A FOR 115VAC; 2.5A FOR 230VAC
INRUSH CURRENT: 65A MAX. FOR 115 VAC; 125A MAX. FOR 230 VAC

DC OUTPUT SPEC

Output Voltage Output Current Min. Output Current Max. Regulation Load Regulation Line Output Ripple & Noise Max.[P-P]
+5V 3 35 ±5% ±50mV 50mV
+12V 2 22 +7/-5% ±50mV 120mV
-5V 0 0.5 ±5% ±50mV 150mV
-12V 0 0.8 ±5% ±50mV 150mV
+3.3V 1.0 28 ±5% ±50mV 50mV
+5VSB 0.1 2.0 ±5% ±50mV 50mV

FEATURES

* TEMPERATURE RANGE: OPERATING -10℃~50℃, STORAGE -40℃~80℃
* HOLD UP TIME: 16 ms MINIMUM AT FULL LOAD & NORMAL INPUT VOLTAGE
* EFFICIENCY: 70% TYPICAL AT 115V, FULL LOAD
* POWER GOOD SIGNAL: ON DELAY 100 ms TO 500 ms, OFF DELAY 1 ms
* OVER POWER PROTECTION: 110%~150%
* OVER VOLTAGE PROTECTION: +3.3V→3.9~4.3V, +5V→5.7~6.5V, +12V→13.6~15.0V
* SHORT CIRCUIT PROTECTION: +5V, +3.3V, +12V, -5V, -12V
* EMI NOISE FILTER: FCC CLASS B, CISPR22 CLASS B
* SAFETY: TO MEET UL, TUV, IEC 60601
* LEAKAGE CURRENT: 0.3mA. MAX. AT NOMINAL VOLTAGE
* COOLING: ONE 80mm DC FAN
* FAN SPEED CONTROL AND NOISE


*THE POWER-SUPPLY IS FOR CHASSIS-ASSEMBLY ONLY AND IS NOT ALLOWED TO BE OPERATED AS STAND-ALONE COMPONENT. FINAL ASSEMBLY HAS TO COMPLY WITH CORRESPONDING EMC- AND SAFETY-REGULATIONS.

 

keywords:zippy,1U Redundant,1U Single,2U Redundant,2U Single,3U Single,Cloud Power,Easy Swap,Medical,Micro Redundant,Mini 1U Single,Mini Redundant,N+1 Redundant,Open Frame,PS2,PS2*2 Redundant,PS2+,Redundant module